Specification
Data is being updated
Description
The R&S®Spectrum Rider FPH offers German engineered performance and ergonomic design at an affordable price. The excellent RF performance can capture weak signals with a DANL as low as –163 dBm. The portable R&S®Spectrum Rider FPH weighs only from 2.5 kg and its battery lasts for 4.5 hours or more. Measure anywhere and anytime at an affordable price.
Features
- Frequency ranges from 5 kHz to 2/3/4 GHz; upgrade via keycode
- Solid RF performance
- Ideal for field use: 8 hour battery life, 2.5 kg (5.5 lb) weight, backlit keypad, fast boot time, non-reflective display, small footprint, ruggedized housing
- Large color display with touch and gesture operation
- Measurement wizard that supports measurement campaigns speeds up measurements and avoids errors
- Features and options for various industries such as aerospace and defense, wireless communications, broadcasting, spectrum regulators and education
- Easy and cost-efficient upgrades of all options via software keycode
Specifications
Frequency range | base unit | 5 kHz to 2 GHz |
with R&S®FPH-B3 option | 5 kHz to 3 GHz | |
with R&S®FPH-B3 and R&S®FPH-B4 option installed | 5 kHz to 4 GHz | |
Frequency resolution | 1 Hz | |
Resolution bandwidth | 1 Hz to 3 MHz in 1/3 sequence | |
Spectral purity SSB phase noise | f = 500 MHz | |
30 kHz | < –88 dBc(1 Hz), typ. –95 dBc(1 Hz) | |
100 kHz | < –98 dBc(1 Hz), typ.–105 dBc(1 Hz) | |
1 MHz | < –118 dBc(1 Hz), typ. –125 dBc(1 Hz) | |
Displayed average noise level | 0 dB RF attenuation, 50 Ω termination, RBW = 100 Hz, VBW = 10 Hz, sample detector, log scaling, tracking generator off, normalized to 1 Hz | |
frequency | preamplifier = off | |
1 MHz to 10 MHz | < –135 dBm, typ. –142 dBm | |
10 MHz to 1 GHz | < –142 dBm, typ. –146 dBm | |
1 GHz to 4 GHz | < –140 dBm, typ. –144 dBm | |
frequency | preamplifier = on | |
1 MHz to 10 MHz | < –150 dBm, typ. –160 dBm | |
10 MHz to 3 GHz | < –158 dBm, typ. –163 dBm | |
3 GHz to 4 GHz | < –156 dBm, typ. –161 dBm | |
Third-order intercept (IP3) | intermodulation-free dynamic range, signal level 2 × –20 dBm, RF attenuation = 0 dB, RF preamplifier = off | |
f = 1 GHz | +7 dBm (meas.) | |
f = 2.4 GHz | +10 dBm (meas.) | |
Total measurement uncertainty | 95 % confidence level, +20 °C to +30 °C, SNR > 16 dB, 0 dB to –50 dB below reference level, RF attenuation auto | |
10 MHz ≤ f ≤ 4 GHz | < 1.25 dB, typ. 0.5 dB | |
Display | ||
Resolution | WVGA, 800 × 480 pixel | |
Battery | lithium ion battery | |
Capacity | R&S®HA-Z306 | 72 Wh |
Voltage | 11.25 V | |
Operating time with new, fully charged battery | R&S®HA-Z306 | 8 h |
Dimensions | W × H × D | 202 mm × 294 mm × 76 mm (8.0 in × 11.6 in × 3 in) |
Weight | 2.5 kg (5.5 lb) |
Accessories